Article Document Academic Article Information Content Entity Journal Article Continuant Continuant Entity Entity Generically Dependent Continuant 2025-05-09T09:57:48 RDF description of Comparison of a 0.55 T dental-dedicated magnetic resonance imaging system with a 1.5 T system in evaluation of the temporomandibular joint regarding subjective image quality assessment and rater agreement - http://repository.healthpartners.com/individual/document-rn46055 <p>OBJECTIVE: We aimed to compare subjective assessment of image quality and rater agreement of magnetic resonance (MR) images of the temporomandibular joint (TMJ) acquired with dental-dedicated 0.55 T and clinical 1.5 T magnetic field strengths. <br>STUDY DESIGN: Six blinded raters evaluated randomized TMJ MR images from 5 healthy participants imaged with 3 pulse sequences on 2 scanners: a 0.55 T system using a custom-built surface coil for dental-dedicated applications and a clinical 1.5 T system using a standard head/neck coil. TMJ anatomy and image quality parameters were assessed using a three-point scale (0 = unacceptable, 1 = acceptable, and 2 = excellent). Data was pooled across participants, pulse sequences, and sides. Kappa coefficients were calculated for intra- and inter-rater agreement. Chi-square tests were applied to explore differences between scanners. Significance was established at P<0.05. <br>RESULTS: Four different raters preferred 0.55 T images for visualizing hard and/or soft tissue anatomy (P � .02). One rater preferred the 0.55 T system regarding contrast. No other differences were detected regarding qualitative ratings. Kappa values for intra-rater agreement ranged from 0.49 to 0.70 for 0.55 T images and 0.42 to 0.71 for 1.5 T images. Inter-rater kappa values ranged from 0.02 to 0.29 for 0.55 T images and 0.02 to 0.23 for 1.5 T images. <br>CONCLUSIONS: Images from the two systems were rated similarly. Further research is needed to assess diagnostic performance in patients.<p> Temporomandibular Joint Disorders Comparison of a 0.55 T dental-dedicated magnetic resonance imaging system with a 1.5 T system in evaluation of the temporomandibular joint regarding subjective image quality assessment and rater agreement 24040 2025-04-30T21:01:35.521-05:00 41660 Dentistry 10.1016/j.oooo.2025.02.011 Oral surgery, oral medicine, oral pathology and oral radiology document-rn46055 public